{"id":767,"date":"2017-06-26T15:29:11","date_gmt":"2017-06-26T14:29:11","guid":{"rendered":"https:\/\/www.groupactionlawyers.co.uk\/blog\/?p=767"},"modified":"2017-06-26T15:29:11","modified_gmt":"2017-06-26T14:29:11","slug":"accu-chek-insight-insulin-pumps-battery-life-issues","status":"publish","type":"post","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","title":{"rendered":"Issues with battery life of the Accu-Chek Insight insulin pumps"},"content":{"rendered":"<p>All Accu-Chek Insight insulin pumps have received new instructions for better device management due to previous complaints about the battery lifetime and the device having unexpected shutdowns.<\/p>\n<p>The first warning for the medical device was issued on the 4<sup>th<\/sup> August 2015 where <em>Roche Diabetes Care<\/em>, the medical manufacturers of the insulin pump, issued advice about the correct battery specifications and appropriate pump settings.<br \/>\n<!--more--><\/p>\n<h2>Warnings issued<\/h2>\n<p>In their urgent field safety notice, Roche Diabetes noted they&#8217;d received feedback from a number of people with diabetes using the Accu-Check insulin pump concerning the battery life on their devices quickly depleting, and the rapid battery depletion was also a cause of the device shutting down spontaneously.<\/p>\n<p>Following a thorough investigation, Roche highlighted that, in these cases, there were low-quality batteries and\/or incorrectly selected battery types for the insulin pump that caused the battery depletion and shutting down of the device.<\/p>\n<p>To rectify this issue, Roche identified Energizer Ultimate Lithium, FR03, 1.5V AAA (FR03) or high-quality alkaline batteries (1.5V AAA (LR03)) as the recommended batteries when using the insulin pump. The medical manufacturer notes that these batteries have been successfully tested with the Accu-Chek Insight insulin pump.<\/p>\n<p>The other recommendations include:<\/p>\n<ul>\n<li>Advising patients not to use carbon zinc or rechargeable nickel cadmium (NiCd\/NiMH) batteries with the Acc-Chek Insight insulin pump and;<\/li>\n<li>Training patients to make sure to always confirm the correct battery type on the insulin pump upon the battery change.<\/li>\n<\/ul>\n<h3>May 2017 alert<\/h3>\n<p>The recommendations from 2015 seemingly weren&#8217;t sufficient enough to tackle the battery life issue as the MHRA issued an updated medical device alert on 2<sup>nd<\/sup> May 2017. In this medical alert, Roche provided new instructions for enhancing the battery life of the insulin pumps. The letter acknowledged that users had been trained on how to change the battery of their insulin pumps, but the manufacturer noted it would like to add additional important measures that could actually improve the battery lifetime as well.<\/p>\n<p>The letter doesn&#8217;t seem to contain anything new from the previous alert. They reiterate that choosing the wrong battery type when changing the batteries, or using low-quality batteries, can lead to rapid depletion of power or result in the system completely shutting off. The only new information that they seem to give is that a rapid voltage drop may prevent the insulin pump from giving out an alarm, which is crucial in notifying users to change their batteries. As a result of the alarm not sounding, there&#8217;s potentially a greater chance that not enough insulin will be delivered to the user.<\/p>\n<p>The risk is heightened during night-time when the user is sleeping.<\/p>\n<h3>Actions<\/h3>\n<p>The MHRA has called for the following actions to be completed by the 6<sup>th<\/sup> June 2017:<\/p>\n<p>Ensure that all users and carers:<\/p>\n<ul>\n<li>Received the field safety notice dated 2<sup>nd<\/sup><\/li>\n<li>Understood the instructions detailed in the notice and follow the advice.<\/li>\n<li>Use the Energizer Ultimate lithium batteries (1.5V AAA\/FR03) as well as follow the steps described in the handling instructions in the notice.<\/li>\n<\/ul>\n<p>Roche has asked for those in receipt of the notice to return the acknowledgement form as the manufacturer hasn&#8217;t received enough responses.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>All Accu-Chek Insight insulin pumps have received new instructions for better device management due to previous complaints about the battery lifetime and the device having unexpected shutdowns. The first warning for the medical device was issued on the 4th August 2015 where Roche Diabetes Care, the medical manufacturers of the insulin pump, issued advice about &hellip; <a href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues\" class=\"more-link\">Continue reading <span class=\"screen-reader-text\">Issues with battery life of the Accu-Chek Insight insulin pumps<\/span><\/a><\/p>\n","protected":false},"author":2,"featured_media":109,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"jetpack_post_was_ever_published":false,"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":false,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2}},"categories":[5],"tags":[],"class_list":["post-767","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-medical"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.5 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Accu-Chek Insight Insulin Pumps Battery Life Issues<\/title>\n<meta name=\"description\" content=\"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues\" \/>\n<meta property=\"og:locale\" content=\"en_GB\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Accu-Chek Insight Insulin Pumps Battery Life Issues\" \/>\n<meta property=\"og:description\" content=\"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues\" \/>\n<meta property=\"og:site_name\" content=\"Group Action Lawyers\" \/>\n<meta property=\"article:published_time\" content=\"2017-06-26T14:29:11+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n\t<meta property=\"og:image:width\" content=\"1200\" \/>\n\t<meta property=\"og:image:height\" content=\"601\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/png\" \/>\n<meta name=\"author\" content=\"Admin\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:image\" content=\"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Admin\" \/>\n\t<meta name=\"twitter:label2\" content=\"Estimated reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"3 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues\"},\"author\":{\"name\":\"Admin\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\"},\"headline\":\"Issues with battery life of the Accu-Chek Insight insulin pumps\",\"datePublished\":\"2017-06-26T14:29:11+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues\"},\"wordCount\":533,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"articleSection\":[\"Medical\"],\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues\",\"name\":\"Accu-Chek Insight Insulin Pumps Battery Life Issues\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"datePublished\":\"2017-06-26T14:29:11+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\"},\"description\":\"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#breadcrumb\"},\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage\",\"url\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"contentUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"width\":725,\"height\":250,\"caption\":\"medical alert\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pumps-battery-life-issues#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Issues with battery life of the Accu-Chek Insight insulin pumps\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/\",\"name\":\"Group Action Lawyers\",\"description\":\"Latest Legal News, Articles &amp; Opinions\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-GB\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\",\"name\":\"Admin\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"caption\":\"Admin\"},\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/author\\\/admin\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Accu-Chek Insight Insulin Pumps Battery Life Issues","description":"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","og_locale":"en_GB","og_type":"article","og_title":"Accu-Chek Insight Insulin Pumps Battery Life Issues","og_description":"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.","og_url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","og_site_name":"Group Action Lawyers","article_published_time":"2017-06-26T14:29:11+00:00","og_image":[{"width":1200,"height":601,"url":"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","type":"image\/png"}],"author":"Admin","twitter_card":"summary_large_image","twitter_image":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","twitter_misc":{"Written by":"Admin","Estimated reading time":"3 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#article","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues"},"author":{"name":"Admin","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781"},"headline":"Issues with battery life of the Accu-Chek Insight insulin pumps","datePublished":"2017-06-26T14:29:11+00:00","mainEntityOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues"},"wordCount":533,"commentCount":0,"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","articleSection":["Medical"],"inLanguage":"en-GB","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","name":"Accu-Chek Insight Insulin Pumps Battery Life Issues","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage"},"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","datePublished":"2017-06-26T14:29:11+00:00","author":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781"},"description":"New instructions have been sent out to users of the Accu-Chek Insight Insulin Pump after complaints were received over the insulin pumps battery life.","breadcrumb":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#breadcrumb"},"inLanguage":"en-GB","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues"]}]},{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#primaryimage","url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","contentUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","width":725,"height":250,"caption":"medical alert"},{"@type":"BreadcrumbList","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.groupactionlawyers.co.uk\/blog"},{"@type":"ListItem","position":2,"name":"Issues with battery life of the Accu-Chek Insight insulin pumps"}]},{"@type":"WebSite","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/","name":"Group Action Lawyers","description":"Latest Legal News, Articles &amp; Opinions","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.groupactionlawyers.co.uk\/blog\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-GB"},{"@type":"Person","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781","name":"Admin","image":{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","caption":"Admin"},"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/author\/admin"}]}},"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","jetpack-related-posts":[{"id":1095,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","url_meta":{"origin":767,"position":0},"title":"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia","author":"Admin","date":"17th November 2017","format":false,"excerpt":"Another issue with insulin devices. These life-saving devices are so important in modern society, so any issue with them is huge. Diabetics and healthcare providers are urged to check Accu-Chek Insight insulin pump display screens regularly as the alarms on the devices may reportedly fail due to an electrical error.\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":107,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","url_meta":{"origin":767,"position":1},"title":"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia","author":"Author","date":"30th August 2016","format":false,"excerpt":"A Medical Device Alert has been issued with the Medicines and Healthcare products Regulatory Agency (MHRA) for the Accu-Chek Insight insulin pump system using NovoRapid PumpCart cartridges which is reportedly prone to leakage. Leakage of the devices, manufactured by Roche Diabetes Care, could lead to serious consequences for users, which\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1626,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/roche-diabetes-care-test-strip-errors-leads-to-recall","url_meta":{"origin":767,"position":2},"title":"Roche Diabetes Care test strip errors leads to recall","author":"Author","date":"4th July 2018","format":false,"excerpt":"Roche Diabetes Care test strip errors has led to a recall of a number of home-use and point-of-care glucose monitoring systems. The Roche Diabetes Care test strip errors are for the following systems: Accu-Chek Aviva; Accu-Chek Performa; and Accu-Chek Inform II. There is a risk of a strip error message\u2026","rel":"","context":"In &quot;Health&quot;","block_context":{"text":"Health","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/health"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1617,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/diabetes-testing-kit-recall","url_meta":{"origin":767,"position":3},"title":"Diabetes testing kit recall","author":"Author","date":"15th June 2018","format":false,"excerpt":"News of a diabetes testing kit recall is said to be \u201curgent\u201d amid fears of lives at risk over false readings. The Accu-Chek Aviva and the Accu-Chek Performa are the diabetes testing kits being recalled that could lead to an overdose or an under-dose of insulin that can lead to\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"unsatisfactory safety information","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1369,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/medtronic-recalls-diabetes-infusion-sets-over-hypoglycemia-risk","url_meta":{"origin":767,"position":4},"title":"Medtronic recalls disposable diabetes infusion sets over risk of causing hypoglycemia","author":"Admin","date":"14th February 2018","format":false,"excerpt":"Medical device manufacturer Medtronic Plc is recalling disposable diabetes infusion sets as a possible defect may trigger excessive doses of insulin to the diabetic user, putting them at risk of hypoglycemia. The recall reportedly affects insulin infusion sets distributed all over the world. The device, connected to an insulin pump,\u2026","rel":"","context":"In &quot;Latest&quot;","block_context":{"text":"Latest","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/latest"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":910,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/medtronic-under-greater-scrutiny-more-problem-products","url_meta":{"origin":767,"position":5},"title":"Medtronic under greater scrutiny as more of its products run into problems","author":"Admin","date":"30th August 2017","format":false,"excerpt":"Medical device maker Medtronic have been making the headlines for all the wrong reasons lately. Priding itself as the \"global leader in medical technology, services, and solutions\", Medtronic have been at the centre of a fair few recalls for defects and risks that have arisen from the use of their\u2026","rel":"","context":"In &quot;Latest&quot;","block_context":{"text":"Latest","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/latest"},"img":{"alt_text":"alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/12\/healthcare-apps.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/12\/healthcare-apps.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/12\/healthcare-apps.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/12\/healthcare-apps.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]}],"jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/767","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/comments?post=767"}],"version-history":[{"count":2,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/767\/revisions"}],"predecessor-version":[{"id":769,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/767\/revisions\/769"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media\/109"}],"wp:attachment":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media?parent=767"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/categories?post=767"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/tags?post=767"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}