{"id":1095,"date":"2017-11-17T14:19:24","date_gmt":"2017-11-17T14:19:24","guid":{"rendered":"https:\/\/www.groupactionlawyers.co.uk\/blog\/?p=1095"},"modified":"2017-11-17T14:19:24","modified_gmt":"2017-11-17T14:19:24","slug":"accu-chek-insight-insulin-pump-alarm-failure-risk","status":"publish","type":"post","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","title":{"rendered":"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia"},"content":{"rendered":"<p>Another issue with insulin devices. These life-saving devices are so important in modern society, so any issue with them is huge.<\/p>\n<p>Diabetics and healthcare providers are urged to check <em>Accu-Chek Insight<\/em> insulin pump display screens regularly as the alarms on the devices may reportedly fail due to an electrical error. Manufactured by <em>Roche Diabetes Care<\/em>, the audible alarm and the vibration alarm may not work properly, therefore failing to notify the user of high blood-sugar levels.<\/p>\n<p><strong>A very serious issue indeed.<\/strong><br \/>\n<!--more--><\/p>\n<h2>About the device<\/h2>\n<p>The <em>Accu-Check Insight<\/em> device, similar in size and look to a pager, is attached to the user through a tube to measure and deliver doses of insulin when the user&#8217;s blood-sugar levels are too high. Some diabetes sufferers struggle produce enough insulin naturally to control their blood-sugar levels, so without an external insulin dose, they could suffer from high blood-sugar; also known as hyperglycaemia.<\/p>\n<p>In serious cases of hyperglycaemia, diabetics could go into a <strong>coma<\/strong> or <strong>suffer severe dehydration<\/strong>.<\/p>\n<h3>Users at risk<\/h3>\n<p>Advertised as a device that lets users get on with their lives, many users of the device may rely on audible or vibration alert, rather than manually check the display for high or low blood-sugar levels. This means if the audible and\/or the vibration alert fails, the user may not realise there is a problem.<\/p>\n<p>Roche&#8217;s Field Safety Notice provides the example of users who rely on the audible or vibrating alarm when sleeping to let them know their blood-sugar level is too high, putting people at risk of hyperglycaemia.<\/p>\n<h3>Spotting the problem<\/h3>\n<p>If there is a problem with the alarms, it should reportedly be detected by the device when the battery or insulin cartridge is changed. When detected, an &#8216;E7&#8217; message should show up on the device display screen. If the user is unable to resolve the E7 error message by using the instructions manual, the user ought to contact Roche by calling the customer helpline. If unresolved, the user may have to ask for a replacement device.<\/p>\n<p>For a medical device that is designed to give the user freedom from having to spend a lot time to check their blood-sugar levels and inject insulin manually when needed, the error is sure to put a lot of people under a lot of stress to check if the device is not working.\u00a0 The defect arguably does anything but &#8220;inspire confidence in diabetes management,&#8221; as labelled on the device website.<\/p>\n<h3>Related product defects<\/h3>\n<p>Only last year, the Medicines and Healthcare products Regulatory Agency issued a press release warning users of the same products to check if they are using the devices correctly as they could easily overdose on insulin and risk hypoglycaemia (low blood sugar). The issue prompted Roche to improve their instructions on how to use the devices.<\/p>\n<p>For those who are concerned about the devices, users are encouraged to speak to a GP or other medical professional about the issue.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Another issue with insulin devices. These life-saving devices are so important in modern society, so any issue with them is huge. Diabetics and healthcare providers are urged to check Accu-Chek Insight insulin pump display screens regularly as the alarms on the devices may reportedly fail due to an electrical error. Manufactured by Roche Diabetes Care, &hellip; <a href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk\" class=\"more-link\">Continue reading <span class=\"screen-reader-text\">Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia<\/span><\/a><\/p>\n","protected":false},"author":2,"featured_media":109,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":true,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2},"jetpack_post_was_ever_published":false},"categories":[5],"tags":[25],"class_list":["post-1095","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-medical","tag-device-alerts"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.5 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Accu-Chek Insight Insulin Pump Alarm Failure Risk<\/title>\n<meta name=\"description\" content=\"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk\" \/>\n<meta property=\"og:locale\" content=\"en_GB\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Accu-Chek Insight Insulin Pump Alarm Failure Risk\" \/>\n<meta property=\"og:description\" content=\"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk\" \/>\n<meta property=\"og:site_name\" content=\"Group Action Lawyers\" \/>\n<meta property=\"article:published_time\" content=\"2017-11-17T14:19:24+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n\t<meta property=\"og:image:width\" content=\"1200\" \/>\n\t<meta property=\"og:image:height\" content=\"601\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/png\" \/>\n<meta name=\"author\" content=\"Admin\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:image\" content=\"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Admin\" \/>\n\t<meta name=\"twitter:label2\" content=\"Estimated reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"2 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk\"},\"author\":{\"name\":\"Admin\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\"},\"headline\":\"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia\",\"datePublished\":\"2017-11-17T14:19:24+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk\"},\"wordCount\":503,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"keywords\":[\"device alerts\"],\"articleSection\":[\"Medical\"],\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk\",\"name\":\"Accu-Chek Insight Insulin Pump Alarm Failure Risk\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"datePublished\":\"2017-11-17T14:19:24+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\"},\"description\":\"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#breadcrumb\"},\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage\",\"url\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"contentUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"width\":725,\"height\":250,\"caption\":\"medical alert\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-alarm-failure-risk#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/\",\"name\":\"Group Action Lawyers\",\"description\":\"Latest Legal News, Articles &amp; Opinions\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-GB\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/3c954adf13d2913933b2069fbd6df781\",\"name\":\"Admin\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g\",\"caption\":\"Admin\"},\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/author\\\/admin\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Accu-Chek Insight Insulin Pump Alarm Failure Risk","description":"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","og_locale":"en_GB","og_type":"article","og_title":"Accu-Chek Insight Insulin Pump Alarm Failure Risk","og_description":"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.","og_url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","og_site_name":"Group Action Lawyers","article_published_time":"2017-11-17T14:19:24+00:00","og_image":[{"width":1200,"height":601,"url":"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","type":"image\/png"}],"author":"Admin","twitter_card":"summary_large_image","twitter_image":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","twitter_misc":{"Written by":"Admin","Estimated reading time":"2 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#article","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk"},"author":{"name":"Admin","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781"},"headline":"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia","datePublished":"2017-11-17T14:19:24+00:00","mainEntityOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk"},"wordCount":503,"commentCount":0,"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","keywords":["device alerts"],"articleSection":["Medical"],"inLanguage":"en-GB","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","name":"Accu-Chek Insight Insulin Pump Alarm Failure Risk","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage"},"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","datePublished":"2017-11-17T14:19:24+00:00","author":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781"},"description":"Diabetics and healthcare providers urged to check Accu-Chek Insight insulin pump display screens as the alarms on devices may fail due to electrical error.","breadcrumb":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#breadcrumb"},"inLanguage":"en-GB","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk"]}]},{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#primaryimage","url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","contentUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","width":725,"height":250,"caption":"medical alert"},{"@type":"BreadcrumbList","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.groupactionlawyers.co.uk\/blog"},{"@type":"ListItem","position":2,"name":"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia"}]},{"@type":"WebSite","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/","name":"Group Action Lawyers","description":"Latest Legal News, Articles &amp; Opinions","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.groupactionlawyers.co.uk\/blog\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-GB"},{"@type":"Person","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/3c954adf13d2913933b2069fbd6df781","name":"Admin","image":{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/80c638f202b7125d0217e22b6b80daa614813aec57ba35fb6741bb7737941f2c?s=96&d=mm&r=g","caption":"Admin"},"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/author\/admin"}]}},"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","jetpack-related-posts":[{"id":767,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","url_meta":{"origin":1095,"position":0},"title":"Issues with battery life of the Accu-Chek Insight insulin pumps","author":"Admin","date":"26th June 2017","format":false,"excerpt":"All Accu-Chek Insight insulin pumps have received new instructions for better device management due to previous complaints about the battery lifetime and the device having unexpected shutdowns. The first warning for the medical device was issued on the 4th August 2015 where Roche Diabetes Care, the medical manufacturers of the\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":107,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","url_meta":{"origin":1095,"position":1},"title":"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia","author":"Author","date":"30th August 2016","format":false,"excerpt":"A Medical Device Alert has been issued with the Medicines and Healthcare products Regulatory Agency (MHRA) for the Accu-Chek Insight insulin pump system using NovoRapid PumpCart cartridges which is reportedly prone to leakage. Leakage of the devices, manufactured by Roche Diabetes Care, could lead to serious consequences for users, which\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1617,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/diabetes-testing-kit-recall","url_meta":{"origin":1095,"position":2},"title":"Diabetes testing kit recall","author":"Author","date":"15th June 2018","format":false,"excerpt":"News of a diabetes testing kit recall is said to be \u201curgent\u201d amid fears of lives at risk over false readings. The Accu-Chek Aviva and the Accu-Chek Performa are the diabetes testing kits being recalled that could lead to an overdose or an under-dose of insulin that can lead to\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"unsatisfactory safety information","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1369,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/medtronic-recalls-diabetes-infusion-sets-over-hypoglycemia-risk","url_meta":{"origin":1095,"position":3},"title":"Medtronic recalls disposable diabetes infusion sets over risk of causing hypoglycemia","author":"Admin","date":"14th February 2018","format":false,"excerpt":"Medical device manufacturer Medtronic Plc is recalling disposable diabetes infusion sets as a possible defect may trigger excessive doses of insulin to the diabetic user, putting them at risk of hypoglycemia. The recall reportedly affects insulin infusion sets distributed all over the world. The device, connected to an insulin pump,\u2026","rel":"","context":"In &quot;Latest&quot;","block_context":{"text":"Latest","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/latest"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":1626,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/roche-diabetes-care-test-strip-errors-leads-to-recall","url_meta":{"origin":1095,"position":4},"title":"Roche Diabetes Care test strip errors leads to recall","author":"Author","date":"4th July 2018","format":false,"excerpt":"Roche Diabetes Care test strip errors has led to a recall of a number of home-use and point-of-care glucose monitoring systems. The Roche Diabetes Care test strip errors are for the following systems: Accu-Chek Aviva; Accu-Chek Performa; and Accu-Chek Inform II. There is a risk of a strip error message\u2026","rel":"","context":"In &quot;Health&quot;","block_context":{"text":"Health","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/health"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1064,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/novopen-insulin-pens-recalled-over-hyperglycaemia-risk","url_meta":{"origin":1095,"position":5},"title":"NovoPen insulin pens recalled over risk of hyperglycaemia","author":"Admin","date":"6th November 2017","format":false,"excerpt":"Certain batches of insulin pens manufactured by Novo Nordisk have been recalled as the cartridge holders may not be strong enough to withstand some household cleaners. The Medicines and Healthcare products Regulatory Agency (MHRA) have issued a medical device alert to notify hospitals and users to return affected batches to\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]}],"jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/1095","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/comments?post=1095"}],"version-history":[{"count":4,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/1095\/revisions"}],"predecessor-version":[{"id":1100,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/1095\/revisions\/1100"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media\/109"}],"wp:attachment":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media?parent=1095"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/categories?post=1095"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/tags?post=1095"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}