{"id":107,"date":"2016-08-30T16:10:00","date_gmt":"2016-08-30T15:10:00","guid":{"rendered":"http:\/\/www.groupactionlawyers.co.uk\/blog\/?p=107"},"modified":"2016-12-21T14:35:40","modified_gmt":"2016-12-21T14:35:40","slug":"accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","status":"publish","type":"post","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","title":{"rendered":"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia"},"content":{"rendered":"<p>A Medical Device Alert has been issued with the Medicines and Healthcare products Regulatory Agency (MHRA) for the Accu-Chek Insight insulin pump system using NovoRapid PumpCart cartridges which is reportedly prone to leakage.<\/p>\n<p>Leakage of the devices, manufactured by Roche Diabetes Care, could lead to serious consequences for users, which could include hyperglycaemia, diabetic ketoacidosis, or even death.<br \/>\n<!--more--><br \/>\nThe leakages may result in an under-delivery of vital insulin for users, which can lead to serious health conditions that can rapidly evolve. This has stemmed for the cartridges being liable to incorrect insertion, and off the back of the alert, manuals are being updated, and users are being asked to take care when replacing cartridges, and check for leakages.<\/p>\n<h2>A serious concern<\/h2>\n<p>Any Medical Device Alert which puts user&#8217;s health at risk is always a serious concern.<\/p>\n<p>An under-delivery of vital insulin for diabetic patients can lead to health complications arising very quickly, so the problem is a serious concern. Healthcare workers responsible for patients, as well as diabetes departments, are also to take action in ensuring that patients are made aware of the problem, and prevent any health complications arising.<\/p>\n<p>Patients are to be made aware of the problem as quickly as possible<\/p>\n<h3>What happens if a patient suffers as a result of the problem?<\/h3>\n<p>The problem appears to be an unforeseen error that has possibly not been addressed in training and manuals as thoroughly as it ought to have been. If there is a large enough lack of understanding for users that this could happen, we have to question whether the delivery system is safe enough, and whether the manuals and guides are robust enough.<\/p>\n<p>In a situation where a device can cause problems which can be entirely preventable through innovation or training, a patient who is adversely affected <strong>may have a right to claim compensation<\/strong> for any suffering caused. When we&#8217;re talking about something a serious as an insulin under-delivery &#8211; which in the worse case scenario can lead to patient death &#8211; it has to be looked at carefully.<\/p>\n<p><strong>Have you been affected?<\/strong><\/p>\n<p>If you have been affected then our <a href=\"\/\">Group Action Lawyers<\/a> would be happy to speak with you to see what rights you may have if you have suffered as a result of the problem. Please don&#8217;t hesitate to contact us for free, friendly, informal, and no obligation advice.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>A Medical Device Alert has been issued with the Medicines and Healthcare products Regulatory Agency (MHRA) for the Accu-Chek Insight insulin pump system using NovoRapid PumpCart cartridges which is reportedly prone to leakage. Leakage of the devices, manufactured by Roche Diabetes Care, could lead to serious consequences for users, which could include hyperglycaemia, diabetic ketoacidosis, &hellip; <a href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\" class=\"more-link\">Continue reading <span class=\"screen-reader-text\">Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia<\/span><\/a><\/p>\n","protected":false},"author":1,"featured_media":109,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":false,"jetpack_social_options":{"image_generator_settings":{"template":"highway","default_image_id":0,"font":"","enabled":false},"version":2},"jetpack_post_was_ever_published":false},"categories":[5],"tags":[4],"class_list":["post-107","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-medical","tag-compensation-claims"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.5 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks<\/title>\n<meta name=\"description\" content=\"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\" \/>\n<meta property=\"og:locale\" content=\"en_GB\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks\" \/>\n<meta property=\"og:description\" content=\"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\" \/>\n<meta property=\"og:site_name\" content=\"Group Action Lawyers\" \/>\n<meta property=\"article:published_time\" content=\"2016-08-30T15:10:00+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2016-12-21T14:35:40+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n\t<meta property=\"og:image:width\" content=\"1200\" \/>\n\t<meta property=\"og:image:height\" content=\"601\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/png\" \/>\n<meta name=\"author\" content=\"Author\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:image\" content=\"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Author\" \/>\n\t<meta name=\"twitter:label2\" content=\"Estimated reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"2 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\"},\"author\":{\"name\":\"Author\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/0203ba576ea8f4e8ec318e9677b2a84b\"},\"headline\":\"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia\",\"datePublished\":\"2016-08-30T15:10:00+00:00\",\"dateModified\":\"2016-12-21T14:35:40+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\"},\"wordCount\":410,\"commentCount\":0,\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"keywords\":[\"compensation claims\"],\"articleSection\":[\"Medical\"],\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\",\"name\":\"Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"datePublished\":\"2016-08-30T15:10:00+00:00\",\"dateModified\":\"2016-12-21T14:35:40+00:00\",\"author\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/0203ba576ea8f4e8ec318e9677b2a84b\"},\"description\":\"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#breadcrumb\"},\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage\",\"url\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"contentUrl\":\"https:\\\/\\\/i0.wp.com\\\/www.groupactionlawyers.co.uk\\\/blog\\\/wp-content\\\/uploads\\\/2016\\\/08\\\/medical_alert.png?fit=725%2C250&ssl=1\",\"width\":725,\"height\":250,\"caption\":\"medical alert\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#website\",\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/\",\"name\":\"Group Action Lawyers\",\"description\":\"Latest Legal News, Articles &amp; Opinions\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-GB\"},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/#\\\/schema\\\/person\\\/0203ba576ea8f4e8ec318e9677b2a84b\",\"name\":\"Author\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g\",\"caption\":\"Author\"},\"url\":\"https:\\\/\\\/www.groupactionlawyers.co.uk\\\/blog\\\/author\\\/matt\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks","description":"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","og_locale":"en_GB","og_type":"article","og_title":"Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks","og_description":"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.","og_url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","og_site_name":"Group Action Lawyers","article_published_time":"2016-08-30T15:10:00+00:00","article_modified_time":"2016-12-21T14:35:40+00:00","og_image":[{"width":1200,"height":601,"url":"https:\/\/i2.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","type":"image\/png"}],"author":"Author","twitter_card":"summary_large_image","twitter_image":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert_large.png?fit=1200%2C601&ssl=1","twitter_misc":{"Written by":"Author","Estimated reading time":"2 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#article","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia"},"author":{"name":"Author","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/0203ba576ea8f4e8ec318e9677b2a84b"},"headline":"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia","datePublished":"2016-08-30T15:10:00+00:00","dateModified":"2016-12-21T14:35:40+00:00","mainEntityOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia"},"wordCount":410,"commentCount":0,"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","keywords":["compensation claims"],"articleSection":["Medical"],"inLanguage":"en-GB","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia","name":"Accu-Chek Insight Insulin Pump System Prone To Insulin Leaks","isPartOf":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage"},"image":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage"},"thumbnailUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","datePublished":"2016-08-30T15:10:00+00:00","dateModified":"2016-12-21T14:35:40+00:00","author":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/0203ba576ea8f4e8ec318e9677b2a84b"},"description":"A medical alert has been issued for Accu-Chek Insight Insulin Pump Systems as they have been found to leak resulting in an under-delivery of vital insulin.","breadcrumb":{"@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#breadcrumb"},"inLanguage":"en-GB","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia"]}]},{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#primaryimage","url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","contentUrl":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","width":725,"height":250,"caption":"medical alert"},{"@type":"BreadcrumbList","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-system-prone-to-insulin-leaks-risk-of-hyperglycaemia#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.groupactionlawyers.co.uk\/blog"},{"@type":"ListItem","position":2,"name":"Accu-Chek Insight insulin pump system prone to insulin leaks &#8211; risk of hyperglycaemia"}]},{"@type":"WebSite","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#website","url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/","name":"Group Action Lawyers","description":"Latest Legal News, Articles &amp; Opinions","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.groupactionlawyers.co.uk\/blog\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-GB"},{"@type":"Person","@id":"https:\/\/www.groupactionlawyers.co.uk\/blog\/#\/schema\/person\/0203ba576ea8f4e8ec318e9677b2a84b","name":"Author","image":{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/secure.gravatar.com\/avatar\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/538ab826f4f2d4a28667f572267412a34ffb2abdd54a9fe5d344f0d525bf201f?s=96&d=mm&r=g","caption":"Author"},"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/author\/matt"}]}},"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1","jetpack-related-posts":[{"id":1095,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pump-alarm-failure-risk","url_meta":{"origin":107,"position":0},"title":"Risk of alarm failure in Accu-Chek Insight insulin pumps could lead to hyperglycaemia","author":"Admin","date":"17th November 2017","format":false,"excerpt":"Another issue with insulin devices. These life-saving devices are so important in modern society, so any issue with them is huge. Diabetics and healthcare providers are urged to check Accu-Chek Insight insulin pump display screens regularly as the alarms on the devices may reportedly fail due to an electrical error.\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":767,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/accu-chek-insight-insulin-pumps-battery-life-issues","url_meta":{"origin":107,"position":1},"title":"Issues with battery life of the Accu-Chek Insight insulin pumps","author":"Admin","date":"26th June 2017","format":false,"excerpt":"All Accu-Chek Insight insulin pumps have received new instructions for better device management due to previous complaints about the battery lifetime and the device having unexpected shutdowns. The first warning for the medical device was issued on the 4th August 2015 where Roche Diabetes Care, the medical manufacturers of the\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1617,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/diabetes-testing-kit-recall","url_meta":{"origin":107,"position":2},"title":"Diabetes testing kit recall","author":"Author","date":"15th June 2018","format":false,"excerpt":"News of a diabetes testing kit recall is said to be \u201curgent\u201d amid fears of lives at risk over false readings. The Accu-Chek Aviva and the Accu-Chek Performa are the diabetes testing kits being recalled that could lead to an overdose or an under-dose of insulin that can lead to\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"unsatisfactory safety information","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=700%2C400 2x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/medical-device-alert.png?fit=1200%2C414&ssl=1&resize=1050%2C600 3x"},"classes":[]},{"id":1064,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/novopen-insulin-pens-recalled-over-hyperglycaemia-risk","url_meta":{"origin":107,"position":3},"title":"NovoPen insulin pens recalled over risk of hyperglycaemia","author":"Admin","date":"6th November 2017","format":false,"excerpt":"Certain batches of insulin pens manufactured by Novo Nordisk have been recalled as the cartridge holders may not be strong enough to withstand some household cleaners. The Medicines and Healthcare products Regulatory Agency (MHRA) have issued a medical device alert to notify hospitals and users to return affected batches to\u2026","rel":"","context":"In &quot;Medical&quot;","block_context":{"text":"Medical","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/medical"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]},{"id":1626,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/roche-diabetes-care-test-strip-errors-leads-to-recall","url_meta":{"origin":107,"position":4},"title":"Roche Diabetes Care test strip errors leads to recall","author":"Author","date":"4th July 2018","format":false,"excerpt":"Roche Diabetes Care test strip errors has led to a recall of a number of home-use and point-of-care glucose monitoring systems. The Roche Diabetes Care test strip errors are for the following systems: Accu-Chek Aviva; Accu-Chek Performa; and Accu-Chek Inform II. There is a risk of a strip error message\u2026","rel":"","context":"In &quot;Health&quot;","block_context":{"text":"Health","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/health"},"img":{"alt_text":"medical alert","src":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=350%2C200 1x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=525%2C300 1.5x, https:\/\/i0.wp.com\/www.groupactionlawyers.co.uk\/blog\/wp-content\/uploads\/2016\/08\/medical_alert.png?fit=725%2C250&ssl=1&resize=700%2C400 2x"},"classes":[]},{"id":1369,"url":"https:\/\/www.groupactionlawyers.co.uk\/blog\/medtronic-recalls-diabetes-infusion-sets-over-hypoglycemia-risk","url_meta":{"origin":107,"position":5},"title":"Medtronic recalls disposable diabetes infusion sets over risk of causing hypoglycemia","author":"Admin","date":"14th February 2018","format":false,"excerpt":"Medical device manufacturer Medtronic Plc is recalling disposable diabetes infusion sets as a possible defect may trigger excessive doses of insulin to the diabetic user, putting them at risk of hypoglycemia. The recall reportedly affects insulin infusion sets distributed all over the world. The device, connected to an insulin pump,\u2026","rel":"","context":"In &quot;Latest&quot;","block_context":{"text":"Latest","link":"https:\/\/www.groupactionlawyers.co.uk\/blog\/category\/latest"},"img":{"alt_text":"","src":"","width":0,"height":0},"classes":[]}],"jetpack_sharing_enabled":true,"_links":{"self":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/107","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/comments?post=107"}],"version-history":[{"count":5,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/107\/revisions"}],"predecessor-version":[{"id":373,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/posts\/107\/revisions\/373"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media\/109"}],"wp:attachment":[{"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/media?parent=107"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/categories?post=107"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.groupactionlawyers.co.uk\/blog\/wp-json\/wp\/v2\/tags?post=107"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}